3.3.16.1: Thin-Layer Chromatography (TLC)

What it is: A separation technique that uses a stationary phase (thin layer of silica or alumina on glass/plastic) and a mobile phase (solvent) to separate and identify compounds in a mixture.

Method

Setup:

  • Draw a baseline in pencil (~1 cm from bottom) on the TLC plate
  • Mark the starting positions of your samples on the baseline (in pencil — ink would dissolve!)
  • Dissolve your organic compound(s) in a suitable solvent (e.g., ethanol)
  • Use a capillary tube to spot the solution onto the baseline (small, concentrated spots)

Running:

  • Pour solvent (eluent) into a chromatography jar to a depth just below the baseline
  • Place the TLC plate in the jar, baseline downward, NOT touching the solvent initially
  • Cover the jar (to create a saturated atmosphere)
  • Solvent travels up the plate by capillary action, carrying the compounds at different rates

Detection:

🔬

Need expert Chemistry tutoring?

I'm Chloe — an Oxford-trained chemist with over 20 years of experience teaching at the UK's top independent schools. I work with a small number of students each year for focused, personalised 1:1 Chemistry tutoring online.

Whether your child is targeting an A*, preparing a Medicine or Oxbridge application, or needs to rebuild their confidence in Chemistry — I'd love to help.

Contact me →
  • When the solvent front reaches near the top (~1 cm from the top), remove the plate and mark the solvent front immediately (in pencil)
  • For colourless compounds: visualise by UV light (if the compounds absorb UV) or by spraying with a locating reagent (e.g., potassium permanganate solution, which turns colourless compounds purple)
  • For coloured compounds: visible without any treatment

Interpretation — Rf Values

Definition: Rf (retardation factor) = distance travelled by compound ÷ distance travelled by solvent front

All measurements taken from the baseline.

Key point: Rf is constant for a given compound in a given solvent system (under the same conditions). This makes it useful for identification.

3.3.16.2: Gas Chromatography (GC)

What it is: A separation technique used for volatile organic compounds. The stationary phase is a liquid on a solid support (or bonded to the wall of a capillary column), and the mobile phase is an inert gas (e.g., helium, nitrogen, argon).

Sample injection: The sample (a solution of the organic compound) is injected into the injector block, which is heated to 150–200°C. The sample vaporises immediately. Carrier gas (mobile phase) carries the vaporised sample into the column.

Separation in the column: The column is a long, thin tube (capillary) coated with a viscous liquid (the stationary phase) or packed with solid support coated with liquid. The column is in an oven, typically heated to 50–250°C depending on the compounds. Compounds with weaker interactions with the stationary phase travel faster and emerge first.

Retention time (tR): The time a compound takes to travel through the column and reach the detector. Each compound has a characteristic retention time under fixed conditions, so retention time can be used for identification.

Peak area: The area under each peak is proportional to the amount (moles) of that compound in the sample.

Chromatography diagram showing separation of compounds

3.3.16.3: High-Performance Liquid Chromatography (HPLC)

What it is: A separation technique for non-volatile or thermally unstable compounds. The stationary phase is a solid (or liquid bonded to a solid), packed into a narrow column. The mobile phase is a liquid (solvent) forced through at high pressure.

Setup: Sample dissolved in mobile phase (solvent), injected into the column under high pressure (50–300 bar). Mobile phase (liquid solvent) flows through the column at a controlled rate. Compounds separate based on their interactions with the stationary phase. Detector (usually UV or refractive index) measures compounds as they elute.

Stationary phase options:

  • Normal phase: Polar stationary phase (e.g., silica). Polar compounds have higher retention.
  • Reverse phase: Non-polar stationary phase (e.g., C18 bonded to silica). Non-polar compounds have higher retention.

Comparison: TLC vs GC vs HPLC

Feature TLC GC HPLC
Sample type Volatile or non-volatile; stable Volatile; thermally stable Non-volatile or thermally unstable
Cost Very cheap Expensive Very expensive
Quantitative? Qualitative only Yes (with standards) Yes (with standards)
Resolution Moderate High Very high

Exam Technique Tips

When calculating Rf values:

  • Always measure from the baseline, not from the start position
  • Measure from the centre of each spot (for TLC) to the solvent front
  • Rf should be between 0 and 1
  • If you get Rf > 1, you’ve measured the solvent front position wrong

Identifying unknown compounds: Compare Rf values of unknown to known standards run on the same plate, same solvent. If Rf values match, the compound is likely the same (but not certain — different compounds can have the same Rf). Use multiple solvents or 2D chromatography to increase certainty.

Gas chromatography problems in exams:

  • Don’t confuse retention time with Rf — GC uses retention time, TLC uses Rf
  • Peak area (or height) is proportional to amount, but only if the response factor is constant
  • You may be given a chromatogram and asked to calculate the percentage composition: (peak area of compound A) ÷ (total peak area) × 100

Complete notes for AQA A Level Chemistry 3.3.16. For past paper practice, search “AQA A Level chemistry chromatography questions” — this topic appears regularly on Paper 1 and Paper 2.

📄 AQA Paper 2 Complete Revision Guide — Free Download

GO BEYOND THE FREE RESOURCES

Free AQA questions are everywhere — but none of them tells you which specification points you have practised and which you have not. The AQA Complete Revision Pack is built the other way round: topic questions and a 40-mark topic test for every specification point of 7405, precisely matched and numbered exactly as AQA numbers it; Paper 3-style questions on all 12 required practicals, which free question banks rarely give you; and predicted papers for the June 2027 exams that students are already searching for. Full mark schemes throughout. £24.99, one-off, instant download.

See the AQA Complete Revision Pack

Leave a Reply

Your email address will not be published. Required fields are marked *